Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis
BACKGROUND: Intensity-modulated radiation therapy (IMRT) may have too many peaks and valleys, making the treatment plan undeliverable. When there are too many fluency differences between adjacent pixels in the X or Y directions, the X and Y smoothing factors are utilized as weighting factors to pena...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997531/ https://www.ncbi.nlm.nih.gov/pubmed/36908492 http://dx.doi.org/10.4103/jmp.jmp_81_22 |
_version_ | 1784903272949088256 |
---|---|
author | Saroj, Dinesh Kumar Yadav, Suresh Paliwal, Neetu |
author_facet | Saroj, Dinesh Kumar Yadav, Suresh Paliwal, Neetu |
author_sort | Saroj, Dinesh Kumar |
collection | PubMed |
description | BACKGROUND: Intensity-modulated radiation therapy (IMRT) may have too many peaks and valleys, making the treatment plan undeliverable. When there are too many fluency differences between adjacent pixels in the X or Y directions, the X and Y smoothing factors are utilized as weighting factors to penalize this behavior. Generally, a high degree of complexity is accompanied by many monitor units (MUs), large number of segments, small sized segments, and complex segment shapes. The degree of plan delivery uncertainty can all increase with a higher detailed fluence map. AIM: This study aims to evaluate the dosimetric effects of various smoothing levels on the planning target volume (PTV) and organs at risk (OARs) for cervix cancer. MATERIALS AND METHODS: IMRT treatment plans were re-optimized by combining several values of the X and Y penalty between 0 and 100. The dose–volume histogram assessed various dosimetric indicators for PTV and OARs. Additionally, gamma passing rates were evaluated and noted as an indicator of the complex treatment plan. RESULTS: At X = 60, Y = 60 fluence map penalty, the conformity index (CI) value reached its highest value of 0.996 ± 0.004. At X = 0, Y = 0, the homogeneity index (HI) was determined to have a maximum value of 0.0628 ± 0.0235. The highest and lowest MU values were 2424.30 ± 471.12 and 1087.80 ± 91.57, respectively, with X = 0, Y = 0 and X = 100, Y = 100. At X = 100, Y = 100, the gamma passing rate reaches its highest value of 99.28% ± 0.44% and minimum value of 85.93% ± 3.87% at X = 0, Y = 0. CONCLUSION: The CI and HI values showed no discernible fluctuation, and the OAR doses were barely affected as smoothing was increased. When the smoothing factor was raised, the number of MUs sharply dropped, and a decrease in the number of segments and higher gamma passing rates were also seen. |
format | Online Article Text |
id | pubmed-9997531 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-99975312023-03-10 Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis Saroj, Dinesh Kumar Yadav, Suresh Paliwal, Neetu J Med Phys Original Article BACKGROUND: Intensity-modulated radiation therapy (IMRT) may have too many peaks and valleys, making the treatment plan undeliverable. When there are too many fluency differences between adjacent pixels in the X or Y directions, the X and Y smoothing factors are utilized as weighting factors to penalize this behavior. Generally, a high degree of complexity is accompanied by many monitor units (MUs), large number of segments, small sized segments, and complex segment shapes. The degree of plan delivery uncertainty can all increase with a higher detailed fluence map. AIM: This study aims to evaluate the dosimetric effects of various smoothing levels on the planning target volume (PTV) and organs at risk (OARs) for cervix cancer. MATERIALS AND METHODS: IMRT treatment plans were re-optimized by combining several values of the X and Y penalty between 0 and 100. The dose–volume histogram assessed various dosimetric indicators for PTV and OARs. Additionally, gamma passing rates were evaluated and noted as an indicator of the complex treatment plan. RESULTS: At X = 60, Y = 60 fluence map penalty, the conformity index (CI) value reached its highest value of 0.996 ± 0.004. At X = 0, Y = 0, the homogeneity index (HI) was determined to have a maximum value of 0.0628 ± 0.0235. The highest and lowest MU values were 2424.30 ± 471.12 and 1087.80 ± 91.57, respectively, with X = 0, Y = 0 and X = 100, Y = 100. At X = 100, Y = 100, the gamma passing rate reaches its highest value of 99.28% ± 0.44% and minimum value of 85.93% ± 3.87% at X = 0, Y = 0. CONCLUSION: The CI and HI values showed no discernible fluctuation, and the OAR doses were barely affected as smoothing was increased. When the smoothing factor was raised, the number of MUs sharply dropped, and a decrease in the number of segments and higher gamma passing rates were also seen. Wolters Kluwer - Medknow 2022 2023-01-10 /pmc/articles/PMC9997531/ /pubmed/36908492 http://dx.doi.org/10.4103/jmp.jmp_81_22 Text en Copyright: © 2023 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Saroj, Dinesh Kumar Yadav, Suresh Paliwal, Neetu Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis |
title | Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis |
title_full | Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis |
title_fullStr | Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis |
title_full_unstemmed | Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis |
title_short | Does Fluence Smoothing Reduce the Complexity of the Intensity-Modulated Radiation Therapy Treatment Plan? A Dosimetric Analysis |
title_sort | does fluence smoothing reduce the complexity of the intensity-modulated radiation therapy treatment plan? a dosimetric analysis |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997531/ https://www.ncbi.nlm.nih.gov/pubmed/36908492 http://dx.doi.org/10.4103/jmp.jmp_81_22 |
work_keys_str_mv | AT sarojdineshkumar doesfluencesmoothingreducethecomplexityoftheintensitymodulatedradiationtherapytreatmentplanadosimetricanalysis AT yadavsuresh doesfluencesmoothingreducethecomplexityoftheintensitymodulatedradiationtherapytreatmentplanadosimetricanalysis AT paliwalneetu doesfluencesmoothingreducethecomplexityoftheintensitymodulatedradiationtherapytreatmentplanadosimetricanalysis |