In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry

BACKGROUND AND PURPOSE: Total skin electron beam therapy (TSEBT) is an important skin-directed radiotherapeutic procedure done in the treatment of cutaneous T-cell lymphomas, namely, mycosis fungoides (MF). This procedure is usually done at larger source-to-surface distances with the patient standin...

Descripción completa

Detalles Bibliográficos
Autores principales: Baba, Misba Hamid, Singh, Benoy Kumar, Wani, Shaqul Qamar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997533/
https://www.ncbi.nlm.nih.gov/pubmed/36908494
http://dx.doi.org/10.4103/jmp.jmp_72_22
_version_ 1784903273450307584
author Baba, Misba Hamid
Singh, Benoy Kumar
Wani, Shaqul Qamar
author_facet Baba, Misba Hamid
Singh, Benoy Kumar
Wani, Shaqul Qamar
author_sort Baba, Misba Hamid
collection PubMed
description BACKGROUND AND PURPOSE: Total skin electron beam therapy (TSEBT) is an important skin-directed radiotherapeutic procedure done in the treatment of cutaneous T-cell lymphomas, namely, mycosis fungoides (MF). This procedure is usually done at larger source-to-surface distances with the patient standing on a rotatory platform. As the patient has to stand in different positions without any rigid immobilization devices, there are chances that the total skin may not get uniformly irradiated which could lead to nonuniform dose distributions. Therefore, all the necessary arrangements should be made to evaluate the dose for different regions of the skin using suitable in vivo dosimeters at the radiotherapy centers offering these treatments. This study aimed to evaluate the consistency between the delivered and planned doses in vivo during TSEBT using Gafchromic EBT3 film dosimetry. MATERIALS AND METHODS: The surface dose for the six MF patients treated for TSEBT at our hospital from 2018 to 2022 was measured and evaluated. 2 cm × 2 cm Gafchromic(®) EBT3 films were used to measure skin dose at reference body positions of clinical interest. All the patients were treated with the modified Stanford technique. The irradiated film strips were analyzed for the dose using the IMRT OmniPro software. The doses at respective positions were expressed as mean dose ± standard deviation and the deviation was calculated as the percentage of the prescribed dose. RESULTS: One hundred and fifty-four Gafchromic(®) EBT3 film strips irradiated on six TSEBT patients showed a maximum dose variation of 2.00 ± 0.14 Gy, in the central body regions. The dose variation in the peripheral areas such as hands and ears was larger. A variation of 2 ± 0.32 Gy was observed on the hands and ears. The uniformity of the dose delivered to maximum body parts was within −7% and +16% for the peripheral areas like hands. The American Association of Physicists in Medicine recommends a dose uniformity of 8% and 4% in the vertical and horizontal patient plane for direct incident beam; however, for oblique incidences like in the modified Stanford technique, the dose variation is about 15%. CONCLUSION: In vivo dosimetry using Gafchromic EBT3 film dosimetry for TSEBT yields objective data to find the under or overdose regions. That can be useful to provide quality treatment, especially when treatments tend to be as complex as TSEBT.
format Online
Article
Text
id pubmed-9997533
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-99975332023-03-10 In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry Baba, Misba Hamid Singh, Benoy Kumar Wani, Shaqul Qamar J Med Phys Original Article BACKGROUND AND PURPOSE: Total skin electron beam therapy (TSEBT) is an important skin-directed radiotherapeutic procedure done in the treatment of cutaneous T-cell lymphomas, namely, mycosis fungoides (MF). This procedure is usually done at larger source-to-surface distances with the patient standing on a rotatory platform. As the patient has to stand in different positions without any rigid immobilization devices, there are chances that the total skin may not get uniformly irradiated which could lead to nonuniform dose distributions. Therefore, all the necessary arrangements should be made to evaluate the dose for different regions of the skin using suitable in vivo dosimeters at the radiotherapy centers offering these treatments. This study aimed to evaluate the consistency between the delivered and planned doses in vivo during TSEBT using Gafchromic EBT3 film dosimetry. MATERIALS AND METHODS: The surface dose for the six MF patients treated for TSEBT at our hospital from 2018 to 2022 was measured and evaluated. 2 cm × 2 cm Gafchromic(®) EBT3 films were used to measure skin dose at reference body positions of clinical interest. All the patients were treated with the modified Stanford technique. The irradiated film strips were analyzed for the dose using the IMRT OmniPro software. The doses at respective positions were expressed as mean dose ± standard deviation and the deviation was calculated as the percentage of the prescribed dose. RESULTS: One hundred and fifty-four Gafchromic(®) EBT3 film strips irradiated on six TSEBT patients showed a maximum dose variation of 2.00 ± 0.14 Gy, in the central body regions. The dose variation in the peripheral areas such as hands and ears was larger. A variation of 2 ± 0.32 Gy was observed on the hands and ears. The uniformity of the dose delivered to maximum body parts was within −7% and +16% for the peripheral areas like hands. The American Association of Physicists in Medicine recommends a dose uniformity of 8% and 4% in the vertical and horizontal patient plane for direct incident beam; however, for oblique incidences like in the modified Stanford technique, the dose variation is about 15%. CONCLUSION: In vivo dosimetry using Gafchromic EBT3 film dosimetry for TSEBT yields objective data to find the under or overdose regions. That can be useful to provide quality treatment, especially when treatments tend to be as complex as TSEBT. Wolters Kluwer - Medknow 2022 2023-01-10 /pmc/articles/PMC9997533/ /pubmed/36908494 http://dx.doi.org/10.4103/jmp.jmp_72_22 Text en Copyright: © 2023 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Baba, Misba Hamid
Singh, Benoy Kumar
Wani, Shaqul Qamar
In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry
title In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry
title_full In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry
title_fullStr In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry
title_full_unstemmed In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry
title_short In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry
title_sort in vivo dosimetry for dose verification of total skin electron beam therapy using gafchromic® ebt3 film dosimetry
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997533/
https://www.ncbi.nlm.nih.gov/pubmed/36908494
http://dx.doi.org/10.4103/jmp.jmp_72_22
work_keys_str_mv AT babamisbahamid invivodosimetryfordoseverificationoftotalskinelectronbeamtherapyusinggafchromicebt3filmdosimetry
AT singhbenoykumar invivodosimetryfordoseverificationoftotalskinelectronbeamtherapyusinggafchromicebt3filmdosimetry
AT wanishaqulqamar invivodosimetryfordoseverificationoftotalskinelectronbeamtherapyusinggafchromicebt3filmdosimetry