A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy
This paper aims to provide guidance and a framework for commissioning tests and tolerances for the ExacTrac Dynamic image-guided and surface-guided radiotherapy (SGRT) system. ExacTrac Dynamic includes a stereoscopic X-ray system, a structured light projector, stereoscopic cameras, thermal camera fo...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997535/ https://www.ncbi.nlm.nih.gov/pubmed/36908493 http://dx.doi.org/10.4103/jmp.jmp_67_22 |
_version_ | 1784903273965158400 |
---|---|
author | Perrett, Ben Ukath, Jaysree Horgan, Emma Noble, Chris Ramachandran, Prabhakar |
author_facet | Perrett, Ben Ukath, Jaysree Horgan, Emma Noble, Chris Ramachandran, Prabhakar |
author_sort | Perrett, Ben |
collection | PubMed |
description | This paper aims to provide guidance and a framework for commissioning tests and tolerances for the ExacTrac Dynamic image-guided and surface-guided radiotherapy (SGRT) system. ExacTrac Dynamic includes a stereoscopic X-ray system, a structured light projector, stereoscopic cameras, thermal camera for SGRT, and has the capability to track breath holds and internal markers. The system provides fast and accurate image guidance and intrafraction guidance for stereotactic radiosurgery and stereotactic ablative radiotherapy. ExacTrac Dynamic was commissioned on a recently installed Elekta Versa HD. Commissioning tests are described including safety, isocenter calibration, dosimetry, image quality, data transfer, SGRT stability, SGRT localization, gating, fusion, implanted markers, breath hold, and end-to-end testing. Custom phantom designs have been implemented for assessment of the deep inspiration breath-hold workflow, the implanted markers workflow, and for gating tests where remote-controlled movement of a phantom is required. Commissioning tests were all found to be in tolerance, with maximum translational and rotational deviations in SGRT of 0.3 mm and 0.4°, respectively, and X-ray image fusion reproducibility standard deviation of 0.08 mm. Tolerances were based on published documents and upon the performance characteristics of the system as specified by the vendor. The unique configuration of ExacTrac Dynamic requires the end user to design commissioning tests that validate the system for use in the clinical implementation adopted in the department. As there are multiple customizable workflows available, tests should be designed around these workflows, and can be ongoing as workflows are progressively introduced into departmental procedures. |
format | Online Article Text |
id | pubmed-9997535 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-99975352023-03-10 A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy Perrett, Ben Ukath, Jaysree Horgan, Emma Noble, Chris Ramachandran, Prabhakar J Med Phys Technical Note This paper aims to provide guidance and a framework for commissioning tests and tolerances for the ExacTrac Dynamic image-guided and surface-guided radiotherapy (SGRT) system. ExacTrac Dynamic includes a stereoscopic X-ray system, a structured light projector, stereoscopic cameras, thermal camera for SGRT, and has the capability to track breath holds and internal markers. The system provides fast and accurate image guidance and intrafraction guidance for stereotactic radiosurgery and stereotactic ablative radiotherapy. ExacTrac Dynamic was commissioned on a recently installed Elekta Versa HD. Commissioning tests are described including safety, isocenter calibration, dosimetry, image quality, data transfer, SGRT stability, SGRT localization, gating, fusion, implanted markers, breath hold, and end-to-end testing. Custom phantom designs have been implemented for assessment of the deep inspiration breath-hold workflow, the implanted markers workflow, and for gating tests where remote-controlled movement of a phantom is required. Commissioning tests were all found to be in tolerance, with maximum translational and rotational deviations in SGRT of 0.3 mm and 0.4°, respectively, and X-ray image fusion reproducibility standard deviation of 0.08 mm. Tolerances were based on published documents and upon the performance characteristics of the system as specified by the vendor. The unique configuration of ExacTrac Dynamic requires the end user to design commissioning tests that validate the system for use in the clinical implementation adopted in the department. As there are multiple customizable workflows available, tests should be designed around these workflows, and can be ongoing as workflows are progressively introduced into departmental procedures. Wolters Kluwer - Medknow 2022 2023-01-10 /pmc/articles/PMC9997535/ /pubmed/36908493 http://dx.doi.org/10.4103/jmp.jmp_67_22 Text en Copyright: © 2023 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Technical Note Perrett, Ben Ukath, Jaysree Horgan, Emma Noble, Chris Ramachandran, Prabhakar A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy |
title | A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy |
title_full | A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy |
title_fullStr | A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy |
title_full_unstemmed | A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy |
title_short | A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy |
title_sort | framework for exactrac dynamic commissioning for stereotactic radiosurgery and stereotactic ablative radiotherapy |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997535/ https://www.ncbi.nlm.nih.gov/pubmed/36908493 http://dx.doi.org/10.4103/jmp.jmp_67_22 |
work_keys_str_mv | AT perrettben aframeworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT ukathjaysree aframeworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT horganemma aframeworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT noblechris aframeworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT ramachandranprabhakar aframeworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT perrettben frameworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT ukathjaysree frameworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT horganemma frameworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT noblechris frameworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy AT ramachandranprabhakar frameworkforexactracdynamiccommissioningforstereotacticradiosurgeryandstereotacticablativeradiotherapy |