Cargando…

Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer

AIM: The goal of this study is to discuss the commissioning and dosimetric parameters achieved during the clinical implementation of an indigenously developed intracavitary (IC) plus interstitial (IS) template for high dose rate (HDR) image-guided brachytherapy (IGBT) in cancer (Ca) cervix. We want...

Descripción completa

Detalles Bibliográficos
Autores principales: Kaliyaperumal, Venkatesan, Banerjee, Susovan, Kataria, Tejinder, Abraham, Susan K, Kamaraj, Dayanithi, Tamilselvan, Singaravelu, Gupta, Deepak, Bisht, Shyam Singh, Narang, Kushal, Shishak, Sorun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997539/
https://www.ncbi.nlm.nih.gov/pubmed/36908497
http://dx.doi.org/10.4103/jmp.jmp_50_22
_version_ 1784903274932994048
author Kaliyaperumal, Venkatesan
Banerjee, Susovan
Kataria, Tejinder
Abraham, Susan K
Kamaraj, Dayanithi
Tamilselvan, Singaravelu
Gupta, Deepak
Bisht, Shyam Singh
Narang, Kushal
Shishak, Sorun
author_facet Kaliyaperumal, Venkatesan
Banerjee, Susovan
Kataria, Tejinder
Abraham, Susan K
Kamaraj, Dayanithi
Tamilselvan, Singaravelu
Gupta, Deepak
Bisht, Shyam Singh
Narang, Kushal
Shishak, Sorun
author_sort Kaliyaperumal, Venkatesan
collection PubMed
description AIM: The goal of this study is to discuss the commissioning and dosimetric parameters achieved during the clinical implementation of an indigenously developed intracavitary (IC) plus interstitial (IS) template for high dose rate (HDR) image-guided brachytherapy (IGBT) in cancer (Ca) cervix. We want to discuss our achieved values of cumulative equi-effective doses (EQD2) for high-risk clinical target volume (HRCTV) and organ at risk (OAR) and compare it with available published results. MATERIALS AND METHODS: Medanta anterior oblique/lateral oblique template has a total of 19 needles including the central tandem. For commissioning the template with needles, the indigenously made acrylic phantom was used. Oblique and straight needles were placed inside the acrylic phantom and a computed tomography (CT) scan was performed. Sixteen patients were treated in HDR IGBT using this template after external-beam radiotherapy. The IGBT plans were evaluated based on EQD2 of target coverage i.e., dose received by 98% (D(98%)_HRCTV), 90% (D(90%)_HRCTV), and 50% (D(50%)_HRCTV) volume of HRCTV, and dose received by 2 cc (D(2cc)) and 0.1 cc (D(0.1cc)) of OAR using linear quadratic (LQ) radiobiological model. RESULTS: The autoradiographic in radiochromic film shows that the distance between the needle tip and the middle of the source position is 6 mm. The mean D98%_HRCTV and D90%_HRCTV was 76.8 Gy (range: 70-87.7 Gy, P < 0.01) and 84.49 Gy (range: 76.6-96.7 Gy, P < 0.01), respectively. Mean EQD2 of D2cc of the bladder, rectum, and sigmoid was 85.6 Gy (range: 77.5-99.6 Gy, P < 0.03), 74.3 Gy (range: 70.9-76.7 Gy, P < 0.05), and 58.3 Gy (range: 50.6-67.9 Gy, P = 0.01), respectively. The mean total reference air kerma at a 1 m distance is 0.489cGy (range: 0.391-0.681cGy). CONCLUSIONS: The indigenously developed template could attain satisfactory standards in terms of set parameters for commissioning and acceptable dose volume relations in our clinical use for treating the advanced Ca cervix patients who need IC + IS type of HDR IGBT. The comparative analysis with contemporary applicators was acceptable.
format Online
Article
Text
id pubmed-9997539
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-99975392023-03-10 Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer Kaliyaperumal, Venkatesan Banerjee, Susovan Kataria, Tejinder Abraham, Susan K Kamaraj, Dayanithi Tamilselvan, Singaravelu Gupta, Deepak Bisht, Shyam Singh Narang, Kushal Shishak, Sorun J Med Phys Original Article AIM: The goal of this study is to discuss the commissioning and dosimetric parameters achieved during the clinical implementation of an indigenously developed intracavitary (IC) plus interstitial (IS) template for high dose rate (HDR) image-guided brachytherapy (IGBT) in cancer (Ca) cervix. We want to discuss our achieved values of cumulative equi-effective doses (EQD2) for high-risk clinical target volume (HRCTV) and organ at risk (OAR) and compare it with available published results. MATERIALS AND METHODS: Medanta anterior oblique/lateral oblique template has a total of 19 needles including the central tandem. For commissioning the template with needles, the indigenously made acrylic phantom was used. Oblique and straight needles were placed inside the acrylic phantom and a computed tomography (CT) scan was performed. Sixteen patients were treated in HDR IGBT using this template after external-beam radiotherapy. The IGBT plans were evaluated based on EQD2 of target coverage i.e., dose received by 98% (D(98%)_HRCTV), 90% (D(90%)_HRCTV), and 50% (D(50%)_HRCTV) volume of HRCTV, and dose received by 2 cc (D(2cc)) and 0.1 cc (D(0.1cc)) of OAR using linear quadratic (LQ) radiobiological model. RESULTS: The autoradiographic in radiochromic film shows that the distance between the needle tip and the middle of the source position is 6 mm. The mean D98%_HRCTV and D90%_HRCTV was 76.8 Gy (range: 70-87.7 Gy, P < 0.01) and 84.49 Gy (range: 76.6-96.7 Gy, P < 0.01), respectively. Mean EQD2 of D2cc of the bladder, rectum, and sigmoid was 85.6 Gy (range: 77.5-99.6 Gy, P < 0.03), 74.3 Gy (range: 70.9-76.7 Gy, P < 0.05), and 58.3 Gy (range: 50.6-67.9 Gy, P = 0.01), respectively. The mean total reference air kerma at a 1 m distance is 0.489cGy (range: 0.391-0.681cGy). CONCLUSIONS: The indigenously developed template could attain satisfactory standards in terms of set parameters for commissioning and acceptable dose volume relations in our clinical use for treating the advanced Ca cervix patients who need IC + IS type of HDR IGBT. The comparative analysis with contemporary applicators was acceptable. Wolters Kluwer - Medknow 2022 2023-01-10 /pmc/articles/PMC9997539/ /pubmed/36908497 http://dx.doi.org/10.4103/jmp.jmp_50_22 Text en Copyright: © 2023 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Kaliyaperumal, Venkatesan
Banerjee, Susovan
Kataria, Tejinder
Abraham, Susan K
Kamaraj, Dayanithi
Tamilselvan, Singaravelu
Gupta, Deepak
Bisht, Shyam Singh
Narang, Kushal
Shishak, Sorun
Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer
title Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer
title_full Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer
title_fullStr Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer
title_full_unstemmed Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer
title_short Commissioning and Dosimetric Results of an Indigenously Developed Intra-Vaginal Template for Interstitial Plus Intracavitary High dose Rate Image-Guided Brachytherapy of Advanced Cervix Cancer
title_sort commissioning and dosimetric results of an indigenously developed intra-vaginal template for interstitial plus intracavitary high dose rate image-guided brachytherapy of advanced cervix cancer
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997539/
https://www.ncbi.nlm.nih.gov/pubmed/36908497
http://dx.doi.org/10.4103/jmp.jmp_50_22
work_keys_str_mv AT kaliyaperumalvenkatesan commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT banerjeesusovan commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT katariatejinder commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT abrahamsusank commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT kamarajdayanithi commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT tamilselvansingaravelu commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT guptadeepak commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT bishtshyamsingh commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT narangkushal commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer
AT shishaksorun commissioninganddosimetricresultsofanindigenouslydevelopedintravaginaltemplateforinterstitialplusintracavitaryhighdoserateimageguidedbrachytherapyofadvancedcervixcancer