Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength
OBJECTIVES: Measurement of skeletal muscle using ultrasonography (US) has received considerable attention as an alternative method of muscle assessment. However, intra- and inter-rater reliability remains controversial. Furthermore, there is no consensus regarding the relationship between muscle ass...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JARM
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9998244/ https://www.ncbi.nlm.nih.gov/pubmed/36909302 http://dx.doi.org/10.2490/prm.20230008 |
_version_ | 1784903432703836160 |
---|---|
author | Ogawa, Masato Matsumoto, Tsuyoshi Harada, Risa Yoshikawa, Ryo Ueda, Yuya Takamiya, Daiki Sakai, Yoshitada |
author_facet | Ogawa, Masato Matsumoto, Tsuyoshi Harada, Risa Yoshikawa, Ryo Ueda, Yuya Takamiya, Daiki Sakai, Yoshitada |
author_sort | Ogawa, Masato |
collection | PubMed |
description | OBJECTIVES: Measurement of skeletal muscle using ultrasonography (US) has received considerable attention as an alternative method of muscle assessment. However, intra- and inter-rater reliability remains controversial. Furthermore, there is no consensus regarding the relationship between muscle assessment using US and muscle mass or physical assessment. We aimed to verify the validity and reliability of muscle measurements using US and its relationships with muscle strength and physical assessment. METHODS: The 22 participants were all healthy men. Quadriceps muscle thickness was measured by US by three different raters. Intraclass correlation coefficient (ICC) was used to assess inter- and intra-rater reliability. The maximum isokinetic strength of the quadriceps and handgrip strength were used as measures of lower and upper muscle strength, respectively. Leg muscle mass was assessed using the leg skeletal muscle index (SMI), measured by body impedance analysis, and calf circumference. RESULTS: The intra-rater reliability was excellent which the ICC(1,1) ranges 0.957-0.993, and ICC(1,3) ranges 0.985-0.998. For inter-rater reliability, the values of 0.904 for ICC(2,1) and 0.966 for ICC(2,3) indicated excellent reliability. Leg SMI was significantly correlated with quadriceps thickness (r=0.36). Maximum isokinetic strength and handgrip strength showed weak but statistically significant correlations with quadriceps thickness (r=0.20, r=0.30, respectively). The correlation between quadriceps thickness and calf circumference was not statistically significant. CONCLUSIONS: Quadriceps muscle assessment using US is a valid and reliable technique for healthy individuals. Quadriceps muscle thickness was significantly positively correlated with upper and lower muscle strength and leg SMI. Muscle thickness assessment could replace full body muscle assessment in clinical settings. |
format | Online Article Text |
id | pubmed-9998244 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | JARM |
record_format | MEDLINE/PubMed |
spelling | pubmed-99982442023-03-11 Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength Ogawa, Masato Matsumoto, Tsuyoshi Harada, Risa Yoshikawa, Ryo Ueda, Yuya Takamiya, Daiki Sakai, Yoshitada Prog Rehabil Med Original Article OBJECTIVES: Measurement of skeletal muscle using ultrasonography (US) has received considerable attention as an alternative method of muscle assessment. However, intra- and inter-rater reliability remains controversial. Furthermore, there is no consensus regarding the relationship between muscle assessment using US and muscle mass or physical assessment. We aimed to verify the validity and reliability of muscle measurements using US and its relationships with muscle strength and physical assessment. METHODS: The 22 participants were all healthy men. Quadriceps muscle thickness was measured by US by three different raters. Intraclass correlation coefficient (ICC) was used to assess inter- and intra-rater reliability. The maximum isokinetic strength of the quadriceps and handgrip strength were used as measures of lower and upper muscle strength, respectively. Leg muscle mass was assessed using the leg skeletal muscle index (SMI), measured by body impedance analysis, and calf circumference. RESULTS: The intra-rater reliability was excellent which the ICC(1,1) ranges 0.957-0.993, and ICC(1,3) ranges 0.985-0.998. For inter-rater reliability, the values of 0.904 for ICC(2,1) and 0.966 for ICC(2,3) indicated excellent reliability. Leg SMI was significantly correlated with quadriceps thickness (r=0.36). Maximum isokinetic strength and handgrip strength showed weak but statistically significant correlations with quadriceps thickness (r=0.20, r=0.30, respectively). The correlation between quadriceps thickness and calf circumference was not statistically significant. CONCLUSIONS: Quadriceps muscle assessment using US is a valid and reliable technique for healthy individuals. Quadriceps muscle thickness was significantly positively correlated with upper and lower muscle strength and leg SMI. Muscle thickness assessment could replace full body muscle assessment in clinical settings. JARM 2023-03-11 /pmc/articles/PMC9998244/ /pubmed/36909302 http://dx.doi.org/10.2490/prm.20230008 Text en 2023 The Japanese Association of Rehabilitation Medicine https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND) 4.0 License. |
spellingShingle | Original Article Ogawa, Masato Matsumoto, Tsuyoshi Harada, Risa Yoshikawa, Ryo Ueda, Yuya Takamiya, Daiki Sakai, Yoshitada Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength |
title | Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength |
title_full | Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength |
title_fullStr | Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength |
title_full_unstemmed | Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength |
title_short | Reliability and Validity of Quadriceps Muscle Thickness Measurements in Ultrasonography: A Comparison with Muscle Mass and Strength |
title_sort | reliability and validity of quadriceps muscle thickness measurements in ultrasonography: a comparison with muscle mass and strength |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9998244/ https://www.ncbi.nlm.nih.gov/pubmed/36909302 http://dx.doi.org/10.2490/prm.20230008 |
work_keys_str_mv | AT ogawamasato reliabilityandvalidityofquadricepsmusclethicknessmeasurementsinultrasonographyacomparisonwithmusclemassandstrength AT matsumototsuyoshi reliabilityandvalidityofquadricepsmusclethicknessmeasurementsinultrasonographyacomparisonwithmusclemassandstrength AT haradarisa reliabilityandvalidityofquadricepsmusclethicknessmeasurementsinultrasonographyacomparisonwithmusclemassandstrength AT yoshikawaryo reliabilityandvalidityofquadricepsmusclethicknessmeasurementsinultrasonographyacomparisonwithmusclemassandstrength AT uedayuya reliabilityandvalidityofquadricepsmusclethicknessmeasurementsinultrasonographyacomparisonwithmusclemassandstrength AT takamiyadaiki reliabilityandvalidityofquadricepsmusclethicknessmeasurementsinultrasonographyacomparisonwithmusclemassandstrength AT sakaiyoshitada reliabilityandvalidityofquadricepsmusclethicknessmeasurementsinultrasonographyacomparisonwithmusclemassandstrength |