Characterisation of an epitaxial GaAs/Medipix2 detector using fluorescence photons
Abstract A high-purity GaAs sensor of 110 μm thickness has been bump bonded to a Medipix2 readout chip. The room temperature spectroscopic response of this device to fluorescence photons in the energy range from 8 to 28 keV is presented and compared to the response of a 300 μm thick Si sensor, also...
Autores principales: | , , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2008
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/j.nima.2008.03.020 http://cds.cern.ch/record/1111393 |
Sumario: | Abstract A high-purity GaAs sensor of 110 μm thickness has been bump bonded to a Medipix2 readout chip. The room temperature spectroscopic response of this device to fluorescence photons in the energy range from 8 to 28 keV is presented and compared to the response of a 300 μm thick Si sensor, also bonded to a Medipix2 chip. The measured photopeak responses are used to calibrate both detectors. The depth of depletion of the GaAs sensor is estimated to be 50 μm at 140 V sensor bias voltage from measurements made using the 8 keV Kα line of a Cu target X-ray tube. |
---|