Cargando…
A passive rem counter based on CR39 SSNTD coupled with a boron converter
A passive neutron rem counter using a CR39 SSNTD coupled with a boron converter has been developed. The rem counter is a polythene sphere with cadmium and lead insets, designed to have a response function proportional to the fluence to ambient dose equivalent conversion coefficients, H*(10)/Φ, for e...
Autores principales: | , , , |
---|---|
Lenguaje: | eng |
Publicado: |
2009
|
Materias: | |
Acceso en línea: | https://dx.doi.org/10.1016/j.radmeas.2009.10.053 http://cds.cern.ch/record/1264083 |