Measuring field multipoles in accelerator magnets with small-apertures by an oscillating wire moved on a circular trajectory

A method based on an oscillating wire for measuring the field quality in accelerator magnets with small apertures of the order of 10 mm is proposed. The wire is positioned step-by-step on the generators of a cylindrical domain inside the magnet ap erture, i.e. its end-points at the stages are moved...

Descripción completa

Detalles Bibliográficos
Autores principales: Arpaia, P, Buzio, M, Garcia Perez, J, Petrone, C, Russenschuck, S, Walckiers, L
Publicado: 2012
Materias:
Acceso en línea:https://dx.doi.org/10.1088/1748-0221/7/05/P05003