Cargando…
Equalisation of the PMT response to charge particles for the LUCID detector of the ATLAS experiment
With a view to the Run II of LHC, 44 Hamamatsu R760 PMTs have been calibrated with a 207Bi source. The working point of each PMT has been xed in order to equalise the PMT gain at a value smaller than $10^{5}$. Studies on the reproducibility and the stability in time of such calibration permitted to...
Autor principal: | |
---|---|
Lenguaje: | eng |
Publicado: |
2014
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/1753814 |