Measurement of the front-end dead-time of the LHCb muon detector and evaluation of its contribution to the muon detection inefficiency
A method is described which allows to deduce the dead-time of the front-end electronics of the LHCb muon detector from a series of measurements performed at different luminosities at a bunch-crossing rate of 20 MHz. The measured values of the dead-time range from 70 ns to 100 ns. These results allow...