The selected device for the ALICE Time-of-Flight array is the Multigap Resistive Plate Chamber (MRPC). We have tested this device at the Gamma Irradiation Facility at CERN to evaluate the rate dependence. We find that the rate capability of the MRPC easily exceeds the 50 Hz/cm sup 2 maximum expected...