Cargando…
Energy Resolution of the CMS ECAL Barrel Super-Module Using MGPA Electronics
A method for extracting the amplitude of a pulse has been developed in order to optimise the energy resolution of a CMS ECAL super-module in the 2003 test beam. The method consists of using a set of weights to extract the amplitude as well as subtract the baseline level event by event. For an incide...
Autores principales: | , |
---|---|
Lenguaje: | eng |
Publicado: |
2004
|
Materias: | |
Acceso en línea: | http://cds.cern.ch/record/787480 |