Cargando…

Energy Resolution of the CMS ECAL Barrel Super-Module Using MGPA Electronics

A method for extracting the amplitude of a pulse has been developed in order to optimise the energy resolution of a CMS ECAL super-module in the 2003 test beam. The method consists of using a set of weights to extract the amplitude as well as subtract the baseline level event by event. For an incide...

Descripción completa

Detalles Bibliográficos
Autores principales: Dewhirst, G, Brunelière, R
Lenguaje:eng
Publicado: 2004
Materias:
Acceso en línea:http://cds.cern.ch/record/787480
Descripción
Sumario:A method for extracting the amplitude of a pulse has been developed in order to optimise the energy resolution of a CMS ECAL super-module in the 2003 test beam. The method consists of using a set of weights to extract the amplitude as well as subtract the baseline level event by event. For an incident beam impact restricted to an area of 4times 4~mm^2 the resolution of the ECAL, which uses the final prototype version of the MGPA electronics, is shown to match the target resolution.