Cargando…

Calibration of a large multi-element neutron counter in the energy range 85-430 MeV

Describes the calibration of a large 60 element neutron counter with a threshold of 2.7 MeV equivalent electron energy. The performance of the counter has been measured in the neutron kinetic energy range 8.5-430 MeV using a neutron beam at the CERN Synchrocyclotron. The results obtained for the eff...

Descripción completa

Detalles Bibliográficos
Autores principales: Strong, J A, Corbett, I F, Esterling, R J, Garvey, J, Green, M G, Harnew, N, Jane, M R, Jobes, M, Mawson, J, McMahon, T, Robertson, A W, Thomas, D H
Lenguaje:eng
Publicado: 1978
Materias:
Acceso en línea:https://dx.doi.org/10.1016/0029-554X(78)90741-3
http://cds.cern.ch/record/879101