Cargando…
Impact of Implementation of BacT/Alert Virtuo on Blood Culture Time to Positivity in Sepsis Patients
Time to positivity (TTP) for blood culture bottles incubated in the BacT/Alert Virtuo instrument was compared to the BacT/Alert 3D. TTP was significantly shorter with the Virtuo (median 16.2 h) than 3D (median 21.1 h; P < 0.001). Switching from 3D to Virtuo significantly improved TTP in this mult...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
American Society for Microbiology
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10101113/ https://www.ncbi.nlm.nih.gov/pubmed/36749050 http://dx.doi.org/10.1128/spectrum.05003-22 |