Cargando…

Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood

The objective of this study was to compare measurements of nonesterified fatty acids (NEFA) between the gold standard diagnostic laboratory method and a handheld NEFA meter (Qucare Pro meter, DFI Co. Ltd.). Three experiments were conducted to study the usability of the meter. In experiment 1 we comp...

Descripción completa

Detalles Bibliográficos
Autores principales: Venjakob, P.L., Bretzinger, L.F., Borchardt, S., Weber, C., Heuwieser, W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10285242/
https://www.ncbi.nlm.nih.gov/pubmed/37360128
http://dx.doi.org/10.3168/jdsc.2022-0269