Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood
The objective of this study was to compare measurements of nonesterified fatty acids (NEFA) between the gold standard diagnostic laboratory method and a handheld NEFA meter (Qucare Pro meter, DFI Co. Ltd.). Three experiments were conducted to study the usability of the meter. In experiment 1 we comp...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10285242/ https://www.ncbi.nlm.nih.gov/pubmed/37360128 http://dx.doi.org/10.3168/jdsc.2022-0269 |
_version_ | 1785061569347977216 |
---|---|
author | Venjakob, P.L. Bretzinger, L.F. Borchardt, S. Weber, C. Heuwieser, W. |
author_facet | Venjakob, P.L. Bretzinger, L.F. Borchardt, S. Weber, C. Heuwieser, W. |
author_sort | Venjakob, P.L. |
collection | PubMed |
description | The objective of this study was to compare measurements of nonesterified fatty acids (NEFA) between the gold standard diagnostic laboratory method and a handheld NEFA meter (Qucare Pro meter, DFI Co. Ltd.). Three experiments were conducted to study the usability of the meter. In experiment 1 we compared results of the meter obtained from measurements in serum and whole blood with results of the gold standard method. Based on the results of experiment 1 we compared the results measured by the meter in whole blood with results obtained from the gold standard method on a larger scale, as we wanted to omit the step of centrifugation with the cow-side test. In experiment 3 we determined the influence of ambient temperature on measurements. Overall, blood samples of 231 cows were collected between 14 and 20 d in milk. The Spearman correlation coefficients (ρ) were calculated and Bland-Altman plots were created to compare the accuracy of the NEFA meter with the gold standard. In addition, in experiment 2 receiver operating characteristic (ROC) curve analyses were performed, to define thresholds for the NEFA meter to detect cows with a NEFA concentration above 0.3, 0.4, and 0.7 mEq/L. In experiment 1, there was a high correlation between NEFA concentrations in whole blood and serum determined by the NEFA meter and the gold standard (ρ = 0.90 for measurement in whole blood; ρ = 0.93 for measurement in serum). In experiment 2, the measurement in whole blood with the NEFA meter was compared with the gold standard. Despite a lower correlation (ρ = 0.79) the ROC curve analyses revealed a high specificity and a moderate sensitivity for lower cut-points (i.e., 0.3 and 0.4 mEq/L). The NEFA meter underestimated especially high concentration of >0.7 mEq/L. Considering thresholds of 0.3, 0.4, and 0.7 mEq/L measured by the gold standard test, sensitivity and specificity were 59.1% and 96.7%, 79.0% and 95.4%, and 86.4% and 95.6%, respectively, when using 0.3, 0.3, and 0.4 mEq/L as thresholds for the NEFA meter. Accuracy was 74.1%, 88.3%, and 93.8% for the 3 thresholds tested. Experiment 3 showed that measurements should be conducted at approximately 21°C (ρ = 0.73) as correlations were poor at 6.2°C and 15.1°C (ρ = 0.18 and 0.22, respectively). |
format | Online Article Text |
id | pubmed-10285242 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-102852422023-06-23 Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood Venjakob, P.L. Bretzinger, L.F. Borchardt, S. Weber, C. Heuwieser, W. JDS Commun Health, Behavior, and Well-being The objective of this study was to compare measurements of nonesterified fatty acids (NEFA) between the gold standard diagnostic laboratory method and a handheld NEFA meter (Qucare Pro meter, DFI Co. Ltd.). Three experiments were conducted to study the usability of the meter. In experiment 1 we compared results of the meter obtained from measurements in serum and whole blood with results of the gold standard method. Based on the results of experiment 1 we compared the results measured by the meter in whole blood with results obtained from the gold standard method on a larger scale, as we wanted to omit the step of centrifugation with the cow-side test. In experiment 3 we determined the influence of ambient temperature on measurements. Overall, blood samples of 231 cows were collected between 14 and 20 d in milk. The Spearman correlation coefficients (ρ) were calculated and Bland-Altman plots were created to compare the accuracy of the NEFA meter with the gold standard. In addition, in experiment 2 receiver operating characteristic (ROC) curve analyses were performed, to define thresholds for the NEFA meter to detect cows with a NEFA concentration above 0.3, 0.4, and 0.7 mEq/L. In experiment 1, there was a high correlation between NEFA concentrations in whole blood and serum determined by the NEFA meter and the gold standard (ρ = 0.90 for measurement in whole blood; ρ = 0.93 for measurement in serum). In experiment 2, the measurement in whole blood with the NEFA meter was compared with the gold standard. Despite a lower correlation (ρ = 0.79) the ROC curve analyses revealed a high specificity and a moderate sensitivity for lower cut-points (i.e., 0.3 and 0.4 mEq/L). The NEFA meter underestimated especially high concentration of >0.7 mEq/L. Considering thresholds of 0.3, 0.4, and 0.7 mEq/L measured by the gold standard test, sensitivity and specificity were 59.1% and 96.7%, 79.0% and 95.4%, and 86.4% and 95.6%, respectively, when using 0.3, 0.3, and 0.4 mEq/L as thresholds for the NEFA meter. Accuracy was 74.1%, 88.3%, and 93.8% for the 3 thresholds tested. Experiment 3 showed that measurements should be conducted at approximately 21°C (ρ = 0.73) as correlations were poor at 6.2°C and 15.1°C (ρ = 0.18 and 0.22, respectively). Elsevier 2023-02-09 /pmc/articles/PMC10285242/ /pubmed/37360128 http://dx.doi.org/10.3168/jdsc.2022-0269 Text en © 2023. https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Health, Behavior, and Well-being Venjakob, P.L. Bretzinger, L.F. Borchardt, S. Weber, C. Heuwieser, W. Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood |
title | Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood |
title_full | Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood |
title_fullStr | Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood |
title_full_unstemmed | Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood |
title_short | Evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood |
title_sort | evaluation of the diagnostic accuracy of a point-of-care device to measure concentrations of nonesterified fatty acids in serum and whole blood |
topic | Health, Behavior, and Well-being |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10285242/ https://www.ncbi.nlm.nih.gov/pubmed/37360128 http://dx.doi.org/10.3168/jdsc.2022-0269 |
work_keys_str_mv | AT venjakobpl evaluationofthediagnosticaccuracyofapointofcaredevicetomeasureconcentrationsofnonesterifiedfattyacidsinserumandwholeblood AT bretzingerlf evaluationofthediagnosticaccuracyofapointofcaredevicetomeasureconcentrationsofnonesterifiedfattyacidsinserumandwholeblood AT borchardts evaluationofthediagnosticaccuracyofapointofcaredevicetomeasureconcentrationsofnonesterifiedfattyacidsinserumandwholeblood AT weberc evaluationofthediagnosticaccuracyofapointofcaredevicetomeasureconcentrationsofnonesterifiedfattyacidsinserumandwholeblood AT heuwieserw evaluationofthediagnosticaccuracyofapointofcaredevicetomeasureconcentrationsofnonesterifiedfattyacidsinserumandwholeblood |