Cargando…
3D-Ring CZT System With New Low- and Medium-Energy Range: Ultrafast Dual-Isotope Lung SPECT/CT Improvement
New 3D-ring CZT systems with low- and medium-energy-range detectors allow for simultaneous dual-isotope lung scintigraphy. We compared 10-, 7-, 5-, and 3-minute acquisitions for (99m)Tc and (81m)Kr performed simultaneously on StarGuide CZT-SPECT/CT and reframed in 50 patients. Ventilation/perfusion...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10317294/ https://www.ncbi.nlm.nih.gov/pubmed/37200461 http://dx.doi.org/10.1097/RLU.0000000000004706 |