A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users
PURPOSE: The robotic system CoFlex for kidney stone removal via flexible ureteroscopy (fURS) by a single surgeon (solo surgery, abbreviated SSU) is introduced. It combines a versatile robotic arm and a commercially available ureteroscope to enable gravity compensation and safety functions like virtu...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10491575/ https://www.ncbi.nlm.nih.gov/pubmed/37032384 http://dx.doi.org/10.1007/s11548-023-02883-5 |
_version_ | 1785104086961487872 |
---|---|
author | Schlenk, Christopher Hagmann, Katharina Steidle, Florian Oliva Maza, Laura Kolb, Alexander Hellings-Kuß, Anja Schöb, Dominik Stefan Klodmann, Julian Miernik, Arkadiusz Albu-Schäffer, Alin |
author_facet | Schlenk, Christopher Hagmann, Katharina Steidle, Florian Oliva Maza, Laura Kolb, Alexander Hellings-Kuß, Anja Schöb, Dominik Stefan Klodmann, Julian Miernik, Arkadiusz Albu-Schäffer, Alin |
author_sort | Schlenk, Christopher |
collection | PubMed |
description | PURPOSE: The robotic system CoFlex for kidney stone removal via flexible ureteroscopy (fURS) by a single surgeon (solo surgery, abbreviated SSU) is introduced. It combines a versatile robotic arm and a commercially available ureteroscope to enable gravity compensation and safety functions like virtual walls. The haptic feedback from the operation site is comparable to manual fURS, as the surgeon actuates all ureteroscope DoF manually. METHODS: The system hardware and software as well as the design of an exploratory user study on the simulator model with non-medical participants and urology surgeons are described. For each user study task both objective measurements (e.g., completion time) and subjective user ratings of workload (using the NASA-TLX) and usability (using the System Usability Scale SUS) were obtained. RESULTS: CoFlex enabled SSU in fURS. The implemented setup procedure resulted in an average added setup time of 341.7 ± 71.6 s, a NASA-TLX value of 25.2 ± 13.3 and a SUS value of 82.9 ± 14.4. The ratio of inspected kidney calyces remained similar for robotic (93.68 %) and manual endoscope guidance (94.74 %), but the NASA-TLX values were higher (58.1 ± 16.0 vs. 48.9 ± 20.1) and the SUS values lower (51.5 ± 19.9 vs. 63.6 ± 15.3) in the robotic scenario. SSU in the fURS procedure increased the overall operation time from 1173.5 ± 355.7 s to 2131.0 ± 338.0 s, but reduced the number of required surgeons from two to one. CONCLUSIONS: The evaluation of CoFlex in a user study covering a complete fURS intervention confirmed the technical feasibility of the concept and its potential to reduce surgeon working time. Future development steps will enhance the system ergonomics, minimize the users’ physical load while interacting with the robot and exploit the logged data from the user study to optimize the current fURS workflow. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s11548-023-02883-5. |
format | Online Article Text |
id | pubmed-10491575 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-104915752023-09-10 A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users Schlenk, Christopher Hagmann, Katharina Steidle, Florian Oliva Maza, Laura Kolb, Alexander Hellings-Kuß, Anja Schöb, Dominik Stefan Klodmann, Julian Miernik, Arkadiusz Albu-Schäffer, Alin Int J Comput Assist Radiol Surg Original Article PURPOSE: The robotic system CoFlex for kidney stone removal via flexible ureteroscopy (fURS) by a single surgeon (solo surgery, abbreviated SSU) is introduced. It combines a versatile robotic arm and a commercially available ureteroscope to enable gravity compensation and safety functions like virtual walls. The haptic feedback from the operation site is comparable to manual fURS, as the surgeon actuates all ureteroscope DoF manually. METHODS: The system hardware and software as well as the design of an exploratory user study on the simulator model with non-medical participants and urology surgeons are described. For each user study task both objective measurements (e.g., completion time) and subjective user ratings of workload (using the NASA-TLX) and usability (using the System Usability Scale SUS) were obtained. RESULTS: CoFlex enabled SSU in fURS. The implemented setup procedure resulted in an average added setup time of 341.7 ± 71.6 s, a NASA-TLX value of 25.2 ± 13.3 and a SUS value of 82.9 ± 14.4. The ratio of inspected kidney calyces remained similar for robotic (93.68 %) and manual endoscope guidance (94.74 %), but the NASA-TLX values were higher (58.1 ± 16.0 vs. 48.9 ± 20.1) and the SUS values lower (51.5 ± 19.9 vs. 63.6 ± 15.3) in the robotic scenario. SSU in the fURS procedure increased the overall operation time from 1173.5 ± 355.7 s to 2131.0 ± 338.0 s, but reduced the number of required surgeons from two to one. CONCLUSIONS: The evaluation of CoFlex in a user study covering a complete fURS intervention confirmed the technical feasibility of the concept and its potential to reduce surgeon working time. Future development steps will enhance the system ergonomics, minimize the users’ physical load while interacting with the robot and exploit the logged data from the user study to optimize the current fURS workflow. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s11548-023-02883-5. Springer International Publishing 2023-04-09 2023 /pmc/articles/PMC10491575/ /pubmed/37032384 http://dx.doi.org/10.1007/s11548-023-02883-5 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Article Schlenk, Christopher Hagmann, Katharina Steidle, Florian Oliva Maza, Laura Kolb, Alexander Hellings-Kuß, Anja Schöb, Dominik Stefan Klodmann, Julian Miernik, Arkadiusz Albu-Schäffer, Alin A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users |
title | A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users |
title_full | A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users |
title_fullStr | A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users |
title_full_unstemmed | A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users |
title_short | A robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users |
title_sort | robotic system for solo surgery in flexible ureteroscopy: development and evaluation with clinical users |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10491575/ https://www.ncbi.nlm.nih.gov/pubmed/37032384 http://dx.doi.org/10.1007/s11548-023-02883-5 |
work_keys_str_mv | AT schlenkchristopher aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT hagmannkatharina aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT steidleflorian aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT olivamazalaura aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT kolbalexander aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT hellingskußanja aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT schobdominikstefan aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT klodmannjulian aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT miernikarkadiusz aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT albuschafferalin aroboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT schlenkchristopher roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT hagmannkatharina roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT steidleflorian roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT olivamazalaura roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT kolbalexander roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT hellingskußanja roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT schobdominikstefan roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT klodmannjulian roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT miernikarkadiusz roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers AT albuschafferalin roboticsystemforsolosurgeryinflexibleureteroscopydevelopmentandevaluationwithclinicalusers |