Cargando…

Oversensing caused by trapped air in the header of subcutaneous cardioverter‐defibrillator generator

In this study, we report two cases with oversensing due to air accumulation in the subcutaneous implantable cardioverter‐defibrillator (S‐ICD) device generator header. If trapped air in the header of the device is suspected, the re‐connection procedure should be considered or the primary vector must...

Descripción completa

Detalles Bibliográficos
Autores principales: Okada, Takuya, Morita, Junji, Kitai, Takayuki, Fujita, Tsutomu, Kondo, Yusuke
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10549796/
https://www.ncbi.nlm.nih.gov/pubmed/37799797
http://dx.doi.org/10.1002/joa3.12912