Oversensing caused by trapped air in the header of subcutaneous cardioverter‐defibrillator generator
In this study, we report two cases with oversensing due to air accumulation in the subcutaneous implantable cardioverter‐defibrillator (S‐ICD) device generator header. If trapped air in the header of the device is suspected, the re‐connection procedure should be considered or the primary vector must...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10549796/ https://www.ncbi.nlm.nih.gov/pubmed/37799797 http://dx.doi.org/10.1002/joa3.12912 |
Sumario: | In this study, we report two cases with oversensing due to air accumulation in the subcutaneous implantable cardioverter‐defibrillator (S‐ICD) device generator header. If trapped air in the header of the device is suspected, the re‐connection procedure should be considered or the primary vector must be used, which might prevent oversensing.[Image: see text] |
---|