Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements

BACKGROUND: Environmental ozone can rapidly degrade cyanine 5 (Cy5), a fluorescent dye commonly used in microarray gene expression studies. Cyanine 3 (Cy3) is much less affected by atmospheric ozone. Degradation of the Cy5 signal relative to the Cy3 signal in 2-color microarrays will adversely reduc...

Descripción completa

Detalles Bibliográficos
Autores principales: Branham, William S, Melvin, Cathy D, Han, Tao, Desai, Varsha G, Moland, Carrie L, Scully, Adam T, Fuscoe, James C
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1800842/
https://www.ncbi.nlm.nih.gov/pubmed/17295919
http://dx.doi.org/10.1186/1472-6750-7-8
_version_ 1782132347946139648
author Branham, William S
Melvin, Cathy D
Han, Tao
Desai, Varsha G
Moland, Carrie L
Scully, Adam T
Fuscoe, James C
author_facet Branham, William S
Melvin, Cathy D
Han, Tao
Desai, Varsha G
Moland, Carrie L
Scully, Adam T
Fuscoe, James C
author_sort Branham, William S
collection PubMed
description BACKGROUND: Environmental ozone can rapidly degrade cyanine 5 (Cy5), a fluorescent dye commonly used in microarray gene expression studies. Cyanine 3 (Cy3) is much less affected by atmospheric ozone. Degradation of the Cy5 signal relative to the Cy3 signal in 2-color microarrays will adversely reduce the Cy5/Cy3 ratio resulting in unreliable microarray data. RESULTS: Ozone in central Arkansas typically ranges between ~22 ppb to ~46 ppb and can be as high as 60–100 ppb depending upon season, meteorological conditions, and time of day. These levels of ozone are common in many areas of the country during the summer. A carbon filter was installed in the laboratory air handling system to reduce ozone levels in the microarray laboratory. In addition, the airflow was balanced to prevent non-filtered air from entering the laboratory. These modifications reduced the ozone within the microarray laboratory to ~2–4 ppb. Data presented here document reductions in Cy5 signal on both in-house produced microarrays and commercial microarrays as a result of exposure to unfiltered air. Comparisons of identically hybridized microarrays exposed to either carbon-filtered or unfiltered air demonstrated the protective effect of carbon-filtration on microarray data as indicated by Cy5 and Cy3 intensities. LOWESS normalization of the data was not able to completely overcome the effect of ozone-induced reduction of Cy5 signal. Experiments were also conducted to examine the effects of high humidity on microarray quality. Modest, but significant, increases in Cy5 and Cy3 signal intensities were observed after 2 or 4 hours at 98–99% humidity compared to 42% humidity. CONCLUSION: Simple installation of carbon filters in the laboratory air handling system resulted in low and consistent ozone levels. This allowed the accurate determination of gene expression by microarray using Cy5 and Cy3 fluorescent dyes.
format Text
id pubmed-1800842
institution National Center for Biotechnology Information
language English
publishDate 2007
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-18008422007-02-17 Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements Branham, William S Melvin, Cathy D Han, Tao Desai, Varsha G Moland, Carrie L Scully, Adam T Fuscoe, James C BMC Biotechnol Methodology Article BACKGROUND: Environmental ozone can rapidly degrade cyanine 5 (Cy5), a fluorescent dye commonly used in microarray gene expression studies. Cyanine 3 (Cy3) is much less affected by atmospheric ozone. Degradation of the Cy5 signal relative to the Cy3 signal in 2-color microarrays will adversely reduce the Cy5/Cy3 ratio resulting in unreliable microarray data. RESULTS: Ozone in central Arkansas typically ranges between ~22 ppb to ~46 ppb and can be as high as 60–100 ppb depending upon season, meteorological conditions, and time of day. These levels of ozone are common in many areas of the country during the summer. A carbon filter was installed in the laboratory air handling system to reduce ozone levels in the microarray laboratory. In addition, the airflow was balanced to prevent non-filtered air from entering the laboratory. These modifications reduced the ozone within the microarray laboratory to ~2–4 ppb. Data presented here document reductions in Cy5 signal on both in-house produced microarrays and commercial microarrays as a result of exposure to unfiltered air. Comparisons of identically hybridized microarrays exposed to either carbon-filtered or unfiltered air demonstrated the protective effect of carbon-filtration on microarray data as indicated by Cy5 and Cy3 intensities. LOWESS normalization of the data was not able to completely overcome the effect of ozone-induced reduction of Cy5 signal. Experiments were also conducted to examine the effects of high humidity on microarray quality. Modest, but significant, increases in Cy5 and Cy3 signal intensities were observed after 2 or 4 hours at 98–99% humidity compared to 42% humidity. CONCLUSION: Simple installation of carbon filters in the laboratory air handling system resulted in low and consistent ozone levels. This allowed the accurate determination of gene expression by microarray using Cy5 and Cy3 fluorescent dyes. BioMed Central 2007-02-12 /pmc/articles/PMC1800842/ /pubmed/17295919 http://dx.doi.org/10.1186/1472-6750-7-8 Text en Copyright © 2007 Branham et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Methodology Article
Branham, William S
Melvin, Cathy D
Han, Tao
Desai, Varsha G
Moland, Carrie L
Scully, Adam T
Fuscoe, James C
Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements
title Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements
title_full Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements
title_fullStr Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements
title_full_unstemmed Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements
title_short Elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements
title_sort elimination of laboratory ozone leads to a dramatic improvement in the reproducibility of microarray gene expression measurements
topic Methodology Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1800842/
https://www.ncbi.nlm.nih.gov/pubmed/17295919
http://dx.doi.org/10.1186/1472-6750-7-8
work_keys_str_mv AT branhamwilliams eliminationoflaboratoryozoneleadstoadramaticimprovementinthereproducibilityofmicroarraygeneexpressionmeasurements
AT melvincathyd eliminationoflaboratoryozoneleadstoadramaticimprovementinthereproducibilityofmicroarraygeneexpressionmeasurements
AT hantao eliminationoflaboratoryozoneleadstoadramaticimprovementinthereproducibilityofmicroarraygeneexpressionmeasurements
AT desaivarshag eliminationoflaboratoryozoneleadstoadramaticimprovementinthereproducibilityofmicroarraygeneexpressionmeasurements
AT molandcarriel eliminationoflaboratoryozoneleadstoadramaticimprovementinthereproducibilityofmicroarraygeneexpressionmeasurements
AT scullyadamt eliminationoflaboratoryozoneleadstoadramaticimprovementinthereproducibilityofmicroarraygeneexpressionmeasurements
AT fuscoejamesc eliminationoflaboratoryozoneleadstoadramaticimprovementinthereproducibilityofmicroarraygeneexpressionmeasurements