Validation of a 3D CT method for measurement of linear wear of acetabular cups: A hip simulator study

BACKGROUND: We evaluated the accuracy and repeatability of a 3D method for polyethylene acetabular cup wear measurements using computed tomography (CT). We propose that the method be used for clinical in vivo assessment of wear in acetabular cups. MATERIAL AND METHODS: Ultra-high molecular weight po...