Cargando…

Non-obvious Problems in Clark Electrode Application at Elevated Temperature and Ways of Their Elimination

Well-known cause of frequent failures of closed oxygen sensors is the appearance of gas bubbles in the electrolyte. The problem is traditionally associated with insufficient sealing of the sensor that is not always true. Study of a typical temperature regime of measurement system based on Clark sens...

Descripción completa

Detalles Bibliográficos
Autores principales: Miniaev, M. V., Belyakova, M. B., Kostiuk, N. V., Leshchenko, D. V., Fedotova, T. A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3747604/
https://www.ncbi.nlm.nih.gov/pubmed/23984188