Cargando…

Measurements of the Weak UV Absorptions of Isoprene and Acetone at 261–275 nm Using Cavity Ringdown Spectroscopy for Evaluation of a Potential Portable Ringdown Breath Analyzer

The weak absorption spectra of isoprene and acetone have been measured in the wavelength range of 261–275 nm using cavity ringdown spectroscopy. The measured absorption cross-sections of isoprene in the wavelength region of 261–266 nm range from 3.65 × 10(−21) cm(2)·molecule(−1) at 261 nm to 1.42 ×...

Descripción completa

Detalles Bibliográficos
Autores principales: Sahay, Peeyush, Scherrer, Susan T., Wang, Chuji
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3758589/
https://www.ncbi.nlm.nih.gov/pubmed/23803787
http://dx.doi.org/10.3390/s130708170