Dose calibrator linearity test: (99m)Tc versus (18)F radioisotopes

OBJECTIVE: The present study was aimed at evaluating the viability of replacing (18)F with (99m)Tc in dose calibrator linearity testing. MATERIALS AND METHODS: The test was performed with sources of (99m)Tc (62 GBq) and (18)F (12 GBq) whose activities were measured up to values lower than 1 MBq. Rat...