Dose calibrator linearity test: (99m)Tc versus (18)F radioisotopes
OBJECTIVE: The present study was aimed at evaluating the viability of replacing (18)F with (99m)Tc in dose calibrator linearity testing. MATERIALS AND METHODS: The test was performed with sources of (99m)Tc (62 GBq) and (18)F (12 GBq) whose activities were measured up to values lower than 1 MBq. Rat...
Autores principales: | Willegaignon, José, Sapienza, Marcelo Tatit, Coura-Filho, George Barberio, Garcez, Alexandre Teles, Alves, Carlos Eduardo Gonzalez Ribeiro, Cardona, Marissa Anabel Rivera, Gutterres, Ricardo Fraga, Buchpiguel, Carlos Alberto |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Colégio Brasileiro de Radiologia e Diagnóstico por
Imagem
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4366026/ https://www.ncbi.nlm.nih.gov/pubmed/25798005 http://dx.doi.org/10.1590/0100-3984.2014.0018 |
Ejemplares similares
-
Dose Calibrator Linearity Testing: Radioisotope (99m)Tc or (18)F? An Alternative for Reducing Costs in Nuclear Medicine Quality Control
por: Willegaignon, José, et al.
Publicado: (2015) -
Radiation safety measures in diagnostic nuclear medicine, based on
the potential radiation dose emitted by radioactive patients
por: Willegaignon, José, et al.
Publicado: (2023) -
Estimating (131)I biokinetics and radiation doses to the red
marrow and whole body in thyroid cancer patients: probe detection versus image
quantification
por: Willegaignon, José, et al.
Publicado: (2016) -
(95g)Tc and (96g)Tc as alternatives to medical radioisotope (99m)Tc
por: Hayakawa, Takehito, et al.
Publicado: (2018) -
Validation of automated image co-registration integrated into
in-house software for voxel-based internal dosimetry on single-photon emission
computed tomography images
por: Leitão, André Luiz Alberti, et al.
Publicado: (2023)