Cargando…

An Optimized High Throughput Clean-Up Method Using Mixed-Mode SPE Plate for the Analysis of Free Arachidonic Acid in Plasma by LC-MS/MS

A high throughput sample preparation method was developed utilizing mixed-mode solid phase extraction (SPE) in 96-well plate format for the determination of free arachidonic acid in plasma by LC-MS/MS. Plasma was mixed with 3% aqueous ammonia and loaded into each well of 96-well plate. After washing...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Wan, Qin, Suzi, Li, Linsen, Chen, Xiaohua, Wang, Qunjie, Wei, Junfu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4383463/
https://www.ncbi.nlm.nih.gov/pubmed/25873969
http://dx.doi.org/10.1155/2015/374819