Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes

Permeation-tube moisture generators are used in industry as calibrated sources of water vapor and carrier gas mixtures. Measurements were made using three permeation-tube moisture generators of the type used in the semiconductor industry. This paper describes repeatability and reproducibility standa...

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Peter H., Kacker, Raghu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4844508/
https://www.ncbi.nlm.nih.gov/pubmed/27413608
http://dx.doi.org/10.6028/jres.108.022
_version_ 1782428786160041984
author Huang, Peter H.
Kacker, Raghu
author_facet Huang, Peter H.
Kacker, Raghu
author_sort Huang, Peter H.
collection PubMed
description Permeation-tube moisture generators are used in industry as calibrated sources of water vapor and carrier gas mixtures. Measurements were made using three permeation-tube moisture generators of the type used in the semiconductor industry. This paper describes repeatability and reproducibility standard deviations in measurement of moisture concentration from such generators. Repeatability refers to measurements within a system and reproducibility refers to measurements between systems. Two independent methods were used to measure the realized concentration of water vapor. The first measurement, the calculated value, was determined using calibrated permeation rate of permeation-tube and flow rate of dry carrier gas. This is the industrial method of evaluating moisture concentration. The second measurement, the measured value, was determined using the low frost-point generator at the National Institute of Standards and Technology (NIST) and a quartz-crystal-micro-balance. Four pairs of independent measurements for each generator and for six nominal levels in the range from10 nL/L to 100 nL/L were made. The characteristic used to quantify repeatability and reproducibility standard deviations in industrial measurements is the calculated value minus the measured value. Repeatability standard deviation ranges from 1 nL/L to 2 nL/L, approximately. Reproducibility standard deviation ranges from 2 nL/L to 8 nL/L, approximately. The documentary ASTM standard E691-99 was used for both data validation and quantification of the repeatability and reproducibility standard deviations.
format Online
Article
Text
id pubmed-4844508
institution National Center for Biotechnology Information
language English
publishDate 2003
publisher [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
record_format MEDLINE/PubMed
spelling pubmed-48445082016-07-13 Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes Huang, Peter H. Kacker, Raghu J Res Natl Inst Stand Technol Article Permeation-tube moisture generators are used in industry as calibrated sources of water vapor and carrier gas mixtures. Measurements were made using three permeation-tube moisture generators of the type used in the semiconductor industry. This paper describes repeatability and reproducibility standard deviations in measurement of moisture concentration from such generators. Repeatability refers to measurements within a system and reproducibility refers to measurements between systems. Two independent methods were used to measure the realized concentration of water vapor. The first measurement, the calculated value, was determined using calibrated permeation rate of permeation-tube and flow rate of dry carrier gas. This is the industrial method of evaluating moisture concentration. The second measurement, the measured value, was determined using the low frost-point generator at the National Institute of Standards and Technology (NIST) and a quartz-crystal-micro-balance. Four pairs of independent measurements for each generator and for six nominal levels in the range from10 nL/L to 100 nL/L were made. The characteristic used to quantify repeatability and reproducibility standard deviations in industrial measurements is the calculated value minus the measured value. Repeatability standard deviation ranges from 1 nL/L to 2 nL/L, approximately. Reproducibility standard deviation ranges from 2 nL/L to 8 nL/L, approximately. The documentary ASTM standard E691-99 was used for both data validation and quantification of the repeatability and reproducibility standard deviations. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2003 2003-06-01 /pmc/articles/PMC4844508/ /pubmed/27413608 http://dx.doi.org/10.6028/jres.108.022 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright.
spellingShingle Article
Huang, Peter H.
Kacker, Raghu
Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes
title Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes
title_full Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes
title_fullStr Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes
title_full_unstemmed Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes
title_short Repeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes
title_sort repeatability and reproducibility standard deviations in the measurement of trace moisture generated using permeation tubes
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4844508/
https://www.ncbi.nlm.nih.gov/pubmed/27413608
http://dx.doi.org/10.6028/jres.108.022
work_keys_str_mv AT huangpeterh repeatabilityandreproducibilitystandarddeviationsinthemeasurementoftracemoisturegeneratedusingpermeationtubes
AT kackerraghu repeatabilityandreproducibilitystandarddeviationsinthemeasurementoftracemoisturegeneratedusingpermeationtubes