Cargando…

A Double-Primary Dead-Weight Tester for Pressures (35–175) kPa in Gage Mode

Primary pressure standards in the atmospheric pressure range are often established using mercury manometers. Less frequently, controlled-clearance dead-weight testers in which one component (normally the piston) has been dimensionally measured have also been used. Recent advances in technology on tw...

Descripción completa

Detalles Bibliográficos
Autores principales: Jain, Kamlesh, Cen, Yueqin, Bowers, Walter J., Schmidt, James W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4844523/
https://www.ncbi.nlm.nih.gov/pubmed/27413591
http://dx.doi.org/10.6028/jres.108.003