A Double-Primary Dead-Weight Tester for Pressures (35–175) kPa in Gage Mode

Primary pressure standards in the atmospheric pressure range are often established using mercury manometers. Less frequently, controlled-clearance dead-weight testers in which one component (normally the piston) has been dimensionally measured have also been used. Recent advances in technology on tw...