Cargando…
Standardization of Rn-222 at the Australian Radiation Laboratory
The standardization of Rn-222 at the Australian Radiation Laboratory involves the calibration of scintillation cells by two methods using standard Ra-226 solutions traceable to the National Institute of Standards and Technology. One of these methods, namely the injection method, involves direct tran...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
1990
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4930044/ https://www.ncbi.nlm.nih.gov/pubmed/28179771 http://dx.doi.org/10.6028/jres.095.020 |