Cargando…

Optic Nerve Head Measurements With Optical Coherence Tomography: A Phantom-Based Study Reveals Differences Among Clinical Devices

PURPOSE: Optical coherence tomography (OCT) can monitor for glaucoma by measuring dimensions of the optic nerve head (ONH) cup and disc. Multiple clinical studies have shown that different OCT devices yield different estimates of retinal dimensions. We developed phantoms mimicking ONH morphology as...

Descripción completa

Detalles Bibliográficos
Autores principales: Agrawal, Anant, Baxi, Jigesh, Calhoun, William, Chen, Chieh-Li, Ishikawa, Hiroshi, Schuman, Joel S., Wollstein, Gadi, Hammer, Daniel X.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4968925/
https://www.ncbi.nlm.nih.gov/pubmed/27409500
http://dx.doi.org/10.1167/iovs.15-18738