The use of a portable three-lead ECG monitor to detect atrial fibrillation in general practice

OBJECTIVE: To evaluate how well an inexpensive portable three-lead ECG monitor PEM identified patients with atrial fibrillation (AF) compared to a normal 12-lead ECG. DESIGN: Cross-sectional method comparison study. SETTING: From April 2014 to February 2015, we included patients coming to the genera...

Descripción completa

Detalles Bibliográficos
Autores principales: Kristensen, Anne N., Jeyam, Brintha, Riahi, Sam, Jensen, Martin B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Taylor & Francis 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5036021/
https://www.ncbi.nlm.nih.gov/pubmed/27409151
http://dx.doi.org/10.1080/02813432.2016.1207151