Practical Uncertainty Limits to the Mass Determination of a Piston-Gage Weight

The mass of a 590-g piston-gage weight was determined with a standard error of 0.057 mg (0.1 ppm). The sources of error are carefully examined. These include air-buoyancy corrections, physically adsorbed surface moisture, and air-convection within the weighing chamber. We conclude that significant i...

Descripción completa

Detalles Bibliográficos
Autores principales: Davis, R. S., Welch, B. E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1988