Cargando…
Practical Uncertainty Limits to the Mass Determination of a Piston-Gage Weight
The mass of a 590-g piston-gage weight was determined with a standard error of 0.057 mg (0.1 ppm). The sources of error are carefully examined. These include air-buoyancy corrections, physically adsorbed surface moisture, and air-convection within the weighing chamber. We conclude that significant i...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
1988
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5178264/ http://dx.doi.org/10.6028/jres.093.149 |