Cargando…

Practical Uncertainty Limits to the Mass Determination of a Piston-Gage Weight

The mass of a 590-g piston-gage weight was determined with a standard error of 0.057 mg (0.1 ppm). The sources of error are carefully examined. These include air-buoyancy corrections, physically adsorbed surface moisture, and air-convection within the weighing chamber. We conclude that significant i...

Descripción completa

Detalles Bibliográficos
Autores principales: Davis, R. S., Welch, B. E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1988
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5178264/
http://dx.doi.org/10.6028/jres.093.149
Descripción
Sumario:The mass of a 590-g piston-gage weight was determined with a standard error of 0.057 mg (0.1 ppm). The sources of error are carefully examined. These include air-buoyancy corrections, physically adsorbed surface moisture, and air-convection within the weighing chamber. We conclude that significant improvement cannot be realized with the conventional weighing techniques available to most piston-gage users.