Influence of patient position and other inherent factors on image quality in two different cone beam computed tomography (CBCT) devices

OBJECTIVES: The aim of this in vitro study was to evaluate how a deviation from the horizontal plane, affects the image quality in two different CBCT-devices. METHODS: A phantom head SK150 (RANDO, The Phantom Laboratory, Salem, NY, USA) was examined in two CBCT-units: Accuitomo 80 and Veraviewepocs...