Determination of [Formula: see text] using MLC‐collimated rectangular fields for absolute dosimetry of the CyberKnife
Traditional CyberKnife (CK) calibration uses TG‐51, which requires [Formula: see text] to be defined using the standard reference condition of 100 cm SSD in a [Formula: see text] field. Since the CK is calibrated using a 6 cm fixed‐aperture collimating cone at 80 cm SAD, the BJR‐25 method is commonl...