Validation of methods for prediction of clinical output levels of active middle ear implants from measurements in human cadaveric ears
Today, the standard method to predict output levels of active middle ear implants (AMEIs) before clinical data are available is stapes vibration measurement in human cadaveric ears, according to ASTM standard F2504-05. Although this procedure is well established, the validity of the predicted output...