Evaluation of gantry rotation overrun in axial CT scanning
The purpose of this study was to develop and evaluate a simple method to assess gantry rotation overrun in a single axial CT scanning. The exposure time in the axial scanning was measured at selected nominal rotation times (400, 700, and 1000 ms) using a solid‐state detector, the RTI's CT dose...