Motion artifacts occurring at the lung/diaphragm interface using 4D CT attenuation correction of 4D PET scans

For PET/CT, fast CT acquisition time can lead to errors in attenuation correction, particularly at the lung/diaphragm interface. Gated 4D PET can reduce motion artifacts, though residual artifacts may persist depending on the CT dataset used for attenuation correction. We performed phantom studies t...