Angular dependence of the output of a kilovoltage X‐ray therapy unit

During the recommissioning of a Philips RT‐250 kilovoltage X‐ray unit, unexpected output variations with tube head rotation (cross‐plane) and tube head tilt (in‐plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0° in‐plane and 0° cross‐plane...

Descripción completa

Detalles Bibliográficos
Autores principales: Goharian, Mehran, Nygren, Ian, Tambasco, Mauro, Barajas, Jose Eduardo Villarreal
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720435/
https://www.ncbi.nlm.nih.gov/pubmed/20717089
http://dx.doi.org/10.1120/jacmp.v11i3.3216