Angular dependence of the output of a kilovoltage X‐ray therapy unit

During the recommissioning of a Philips RT‐250 kilovoltage X‐ray unit, unexpected output variations with tube head rotation (cross‐plane) and tube head tilt (in‐plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0° in‐plane and 0° cross‐plane...

Descripción completa

Detalles Bibliográficos
Autores principales: Goharian, Mehran, Nygren, Ian, Tambasco, Mauro, Barajas, Jose Eduardo Villarreal
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720435/
https://www.ncbi.nlm.nih.gov/pubmed/20717089
http://dx.doi.org/10.1120/jacmp.v11i3.3216
_version_ 1783284648467496960
author Goharian, Mehran
Nygren, Ian
Tambasco, Mauro
Barajas, Jose Eduardo Villarreal
author_facet Goharian, Mehran
Nygren, Ian
Tambasco, Mauro
Barajas, Jose Eduardo Villarreal
author_sort Goharian, Mehran
collection PubMed
description During the recommissioning of a Philips RT‐250 kilovoltage X‐ray unit, unexpected output variations with tube head rotation (cross‐plane) and tube head tilt (in‐plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0° in‐plane and 0° cross‐plane) over the possible range of angles of in‐plane rotation for 75 kVp (half‐value layer, [Formula: see text]). A less pronounced but noticeable output change (with respect to the neutral position) was observed for cross‐plane rotation reaching 2% for the 225 kVp beam ([Formula: see text]). This output variation was observed while manually adjusting the current to maintain constancy according to the current meter gauge. In order to address the observed output dependence with head orientation, the dose rate monitor chamber of the kilovoltage unit was calibrated to monitor the beam output in real time. The dose rate was manually adjusted to maintain a constant dose rate (in r/min) as displayed on the r/min gauge. This approach resulted in maintaining beam output for the 75 kVp and the 225 kVp beams within [Formula: see text] for the in‐plane angle variation and [Formula: see text] for the cross‐plane angle variation. A daily output check that includes ion chamber‐based measurements at the neutral position and an in‐plane angle of 45° has been implemented using the constant dose rate approach to monitor the stability of the X‐ray beams. As a result of the output variations with in/cross‐plane rotation, the quality control (QC) procedures that are typically used for clinical setup have been modified to test the stability of the beams under the non‐neutral positioning of the X‐ray tube. PACS number: 87.56.Fc
format Online
Article
Text
id pubmed-5720435
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57204352018-04-02 Angular dependence of the output of a kilovoltage X‐ray therapy unit Goharian, Mehran Nygren, Ian Tambasco, Mauro Barajas, Jose Eduardo Villarreal J Appl Clin Med Phys Technical Notes During the recommissioning of a Philips RT‐250 kilovoltage X‐ray unit, unexpected output variations with tube head rotation (cross‐plane) and tube head tilt (in‐plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0° in‐plane and 0° cross‐plane) over the possible range of angles of in‐plane rotation for 75 kVp (half‐value layer, [Formula: see text]). A less pronounced but noticeable output change (with respect to the neutral position) was observed for cross‐plane rotation reaching 2% for the 225 kVp beam ([Formula: see text]). This output variation was observed while manually adjusting the current to maintain constancy according to the current meter gauge. In order to address the observed output dependence with head orientation, the dose rate monitor chamber of the kilovoltage unit was calibrated to monitor the beam output in real time. The dose rate was manually adjusted to maintain a constant dose rate (in r/min) as displayed on the r/min gauge. This approach resulted in maintaining beam output for the 75 kVp and the 225 kVp beams within [Formula: see text] for the in‐plane angle variation and [Formula: see text] for the cross‐plane angle variation. A daily output check that includes ion chamber‐based measurements at the neutral position and an in‐plane angle of 45° has been implemented using the constant dose rate approach to monitor the stability of the X‐ray beams. As a result of the output variations with in/cross‐plane rotation, the quality control (QC) procedures that are typically used for clinical setup have been modified to test the stability of the beams under the non‐neutral positioning of the X‐ray tube. PACS number: 87.56.Fc John Wiley and Sons Inc. 2010-06-09 /pmc/articles/PMC5720435/ /pubmed/20717089 http://dx.doi.org/10.1120/jacmp.v11i3.3216 Text en © 2010 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Goharian, Mehran
Nygren, Ian
Tambasco, Mauro
Barajas, Jose Eduardo Villarreal
Angular dependence of the output of a kilovoltage X‐ray therapy unit
title Angular dependence of the output of a kilovoltage X‐ray therapy unit
title_full Angular dependence of the output of a kilovoltage X‐ray therapy unit
title_fullStr Angular dependence of the output of a kilovoltage X‐ray therapy unit
title_full_unstemmed Angular dependence of the output of a kilovoltage X‐ray therapy unit
title_short Angular dependence of the output of a kilovoltage X‐ray therapy unit
title_sort angular dependence of the output of a kilovoltage x‐ray therapy unit
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720435/
https://www.ncbi.nlm.nih.gov/pubmed/20717089
http://dx.doi.org/10.1120/jacmp.v11i3.3216
work_keys_str_mv AT goharianmehran angulardependenceoftheoutputofakilovoltagexraytherapyunit
AT nygrenian angulardependenceoftheoutputofakilovoltagexraytherapyunit
AT tambascomauro angulardependenceoftheoutputofakilovoltagexraytherapyunit
AT barajasjoseeduardovillarreal angulardependenceoftheoutputofakilovoltagexraytherapyunit