Evaluation of new data processing algorithms for planar gated ventriculography (MUGA)
Before implementing one of two new LVEF radionuclide gated ventriculogram (MUGA) systems, the results from 312 consecutive parallel patient studies were evaluated. Each gamma‐camera acquisition was simultaneously processed by semi‐automatic Medasys Pinnacle and by fully automatic and semiautomatic P...