Cargando…
Evaluation of new data processing algorithms for planar gated ventriculography (MUGA)
Before implementing one of two new LVEF radionuclide gated ventriculogram (MUGA) systems, the results from 312 consecutive parallel patient studies were evaluated. Each gamma‐camera acquisition was simultaneously processed by semi‐automatic Medasys Pinnacle and by fully automatic and semiautomatic P...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720542/ https://www.ncbi.nlm.nih.gov/pubmed/19692979 http://dx.doi.org/10.1120/jacmp.v10i3.2977 |
_version_ | 1783284670812651520 |
---|---|
author | Fair, Joanna R. Heintz, Philip H. Telepak, Robert J. |
author_facet | Fair, Joanna R. Heintz, Philip H. Telepak, Robert J. |
author_sort | Fair, Joanna R. |
collection | PubMed |
description | Before implementing one of two new LVEF radionuclide gated ventriculogram (MUGA) systems, the results from 312 consecutive parallel patient studies were evaluated. Each gamma‐camera acquisition was simultaneously processed by semi‐automatic Medasys Pinnacle and by fully automatic and semiautomatic Philips nuclear medicine computer systems. The Philips systems yielded LVEF results within [Formula: see text] percentage points of the Medasys system in fewer than half of the studies. The remaining values were higher or lower than those from the long‐used Medasys system. These differences might have changed cancer patient chemotherapy clinical decisions. As a result, our institution elected not to implement either new system. PACS: 87.57.U‐ Nuclear medicine imaging |
format | Online Article Text |
id | pubmed-5720542 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2009 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57205422018-04-02 Evaluation of new data processing algorithms for planar gated ventriculography (MUGA) Fair, Joanna R. Heintz, Philip H. Telepak, Robert J. J Appl Clin Med Phys Medical Imaging Before implementing one of two new LVEF radionuclide gated ventriculogram (MUGA) systems, the results from 312 consecutive parallel patient studies were evaluated. Each gamma‐camera acquisition was simultaneously processed by semi‐automatic Medasys Pinnacle and by fully automatic and semiautomatic Philips nuclear medicine computer systems. The Philips systems yielded LVEF results within [Formula: see text] percentage points of the Medasys system in fewer than half of the studies. The remaining values were higher or lower than those from the long‐used Medasys system. These differences might have changed cancer patient chemotherapy clinical decisions. As a result, our institution elected not to implement either new system. PACS: 87.57.U‐ Nuclear medicine imaging John Wiley and Sons Inc. 2009-07-08 /pmc/articles/PMC5720542/ /pubmed/19692979 http://dx.doi.org/10.1120/jacmp.v10i3.2977 Text en © 2009 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Medical Imaging Fair, Joanna R. Heintz, Philip H. Telepak, Robert J. Evaluation of new data processing algorithms for planar gated ventriculography (MUGA) |
title | Evaluation of new data processing algorithms for planar gated ventriculography (MUGA) |
title_full | Evaluation of new data processing algorithms for planar gated ventriculography (MUGA) |
title_fullStr | Evaluation of new data processing algorithms for planar gated ventriculography (MUGA) |
title_full_unstemmed | Evaluation of new data processing algorithms for planar gated ventriculography (MUGA) |
title_short | Evaluation of new data processing algorithms for planar gated ventriculography (MUGA) |
title_sort | evaluation of new data processing algorithms for planar gated ventriculography (muga) |
topic | Medical Imaging |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5720542/ https://www.ncbi.nlm.nih.gov/pubmed/19692979 http://dx.doi.org/10.1120/jacmp.v10i3.2977 |
work_keys_str_mv | AT fairjoannar evaluationofnewdataprocessingalgorithmsforplanargatedventriculographymuga AT heintzphiliph evaluationofnewdataprocessingalgorithmsforplanargatedventriculographymuga AT telepakrobertj evaluationofnewdataprocessingalgorithmsforplanargatedventriculographymuga |