Cargando…
Comparison between measured and calculated dynamic wedge dose distributions using the anisotropic analytic algorithm and pencil‐beam convolution
We used the two available calculation algorithms of the Varian Eclipse 7.3 three‐dimensional (3D) treatment planning system (TPS), the anisotropic analytic algorithm (AAA) and pencil‐beam convolution (PBC), to compare measured and calculated two‐dimensional enhanced dynamic wedge (2D EDW) dose distr...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2007
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722401/ https://www.ncbi.nlm.nih.gov/pubmed/17592453 http://dx.doi.org/10.1120/jacmp.v8i1.2370 |