Cargando…

A device for precision positioning and alignment of room lasers to diminish their contribution to patient setup errors

This report presents an analysis of patient setup errors resulting from inaccurately positioned wall lasers. It suggests that laser beams should agree within 0.2 degree or better with the machine axes that they are delineating. For typical simulator and treatment rooms having wall‐to‐isocenter dista...

Descripción completa

Detalles Bibliográficos
Autores principales: Brezovich, Ivan A., Jordan, Stephen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722625/
https://www.ncbi.nlm.nih.gov/pubmed/18449147
http://dx.doi.org/10.1120/jacmp.v8i4.2398