Verification of dynamic and segmental IMRT delivery by dynamic log file analysis
A program has been developed to evaluate the delivered fluence of step‐and‐shoot segmental and sliding window dynamic multileaf collimator (MLC) fields. To automate these checks, a number of tools have been developed using data available from the dynamic log files that can be created each time a dyn...