Evaluation of MLC leaf positioning accuracy for static and dynamic IMRT treatments using DAVID in vivo dosimetric system*
Accuracy and precision of leaf positioning in multileaf collimators (MLCs) are significant factors for the accuracy of IMRT treatments. This study aimed to investigate the accuracy and repeatability of the MLC leaf positioning via the DAVID in vivo dosimetric system for dynamic and static MLC system...