Cargando…

Impact of a commercially available model-based dose calculation algorithm on treatment planning of high-dose-rate brachytherapy in patients with cervical cancer

We evaluated the impact of model-based dose calculation algorithms (MBDCAs) on high-dose-rate brachytherapy (HDR-BT) treatment planning for patients with cervical cancer. Seven patients with cervical cancer treated using HDR-BT were studied. Tandem and ovoid applicators were used in four patients, a...

Descripción completa

Detalles Bibliográficos
Autores principales: Abe, Kota, Kadoya, Noriyuki, Sato, Shinya, Hashimoto, Shimpei, Nakajima, Yujiro, Miyasaka, Yuya, Ito, Kengo, Umezawa, Rei, Yamamoto, Takaya, Takahashi, Noriyoshi, Takeda, Ken, Jingu, Keiichi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5951107/
https://www.ncbi.nlm.nih.gov/pubmed/29378024
http://dx.doi.org/10.1093/jrr/rrx081